Pneumatic High Pressure Test Pump
Pressure range:95% vacuum to 16Mpa ( 160 bar ); -0 .095 to 16Mpa
Temperature:0 to 50℃
Humidity:< 85%RH
Adjustment resolution:0.1mbar
Overpressure Safety Limit: 180 bar
Pressure media:Air
Net Weight: 5.76 k g
Dimensions: 3 8 7 × 250 × 160mm (long × Wide × High)
Standard Table Interface:M20×1.5(customized:1/4NPT,1/2NPT,G1/4,G1/2)

1
Pressure/vacuum selector
(Positive pressure: pull it out; Vacuum: push it in)
Do not switch while unit is under any pressure
2
Isolation valve/ shut-off valve
isolates the calibration volume from the hand pump and
check valve, ensure measurement stability
3
Standard Table Interface
Used to connect standard pressure gauges( M 20× 1.5)
4
Pressure handle
Upward lifting and downward pressing
5
Maintenance cleaning port
Filter oil and impurities into the pressure system
6
Fine adjust handle
Accurately adjust the applied pressure. Turn Clockwise to increase pressure
7
Vent valve/relief valve
Loosen and release the pressure inside the pump, tighten the valve to pressurize
8
Test interface/Quick connector
Connect the calibrated pressure instrument interface(M20 ×1. 5)
9
Pressurized handle locking pin
Locking the pressure handle (easy to carry)